PMC PROCEDURE (continued)
NM2B
6.23.17
After the first breath is detected, does the APNEA-VOLUME in the
Warning column and the MINUTE VOLUME LO in the Caution column
deactivate? ___ (Y)
6.23.18
Adjust the low alarm limit above the indicated minute volume.
6.23.19
Does the MINUTE VOLUME LO message appear in the Caution
column? ___ (Y)
6.23.20
Adjust the low alarm limit below the indicated minute volume.
6.23.21
Does the MINUTE VOLUME LO message in the Caution column
deactivate? ___ (Y)
6.23.22
Increase the tidal volume to 1000 ml and the frequency to 10 BPM.
6.23.23
Press the O
2
Flush momentarily to inflate the bellows.
( )
6.23.24
Are the tidal and minute volumes on the Spiromed and on the test
volumeter within 20% of each other? ___ (Y)
6.23.25
Create a reverse flow by loosening the expiratory valve dome.
( )
6.23.26
Each time a reverse flow greater than 20 ml is detected, does the
REVERSE FLOW message appear in the Advisory column? ___ (Y)
6.23.27
Tighten the expiratory valve dome.
6.23.28
Disconnect the Spiromed sensor from the VOLUME SENSOR interface.
6.23.29
Do the VOL SENSOR DISC and VOL ALARMS OFF messages appear
in the Advisory column? ___ (Y)
6.23.30
Connect the Spiromed sensor to the Volume Sensor interface and verify
that the alarms clear.
Rev. Y
6-52
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...