6.18 AUXILIARY OXYGEN FLOWMETER - if applicable
6.18.1
Close the auxiliary oxygen flowmeter flow control valve.
6.18.2
Connect a cm H
2
O pressure manometer to the outlet.
6.18.3
Is there an increase in pressure? ___ (N)
6.18.4
Apply a pressure of 50 cm H
2
O to the manometer.
6.18.5
After 30 seconds, what is the pressure on the manometer? ___(>40 cm
H
2
O)
6.18.6
Remove the gauge and test fixture.
6.18.7
Is it possible to adjust the flow over the full range of the auxiliary
oxygen flowmeter? ___ (Y)
6.18.8
Set the flow rate to 5 l/min.
6.18.9
Hold the sensor from a calibrated O
2
Med at the auxiliary oxygen
flowmeter outlet.
6.18.10
After the value stabilizes, what is the oxygen concentration? ___ % (97-
100)
6.18.11
Remove the O
2
Med sensor from the auxiliary oxygen flowmeter, and
insert it into the inspiratory valve dome adapter.
6.18.12
Close the flowmeter flow control valve.