
NM2B
DIAGNOSTICS (continued)
2.1.2
Calibration Menu Screen
Pressing the CALIBRATION key on the Service Menu Screen brings up
the Calibration Menu Screen. This menu allows the TSR to perform
calibration for the Oxygen Analyzer sensor and the Breathing Pressure
Monitor. Figure 2-5 shows a typical Calibration Menu Screen. The four-
digit numbers in the left column show the current values, while the
numbers in the right column show the values stored from the previous
calibration.
For a valid zero calibration of the Oxygen Analyzer sensor, the two
offset readings should be between 125 and 145, and the difference
between Cell A and Cell B should be no greater than 8. Pressing the
O2MED key stores the values, and the message OXYGEN OFFSET
READINGS STORED appears on the screen. Refer to Section 5 of this
manual for the complete calibration procedure.
For a valid calibration of the Breathing Pressure Monitor, the pressure
span reading at 60 cm H
2
O should be between 465 and 519. Pressing
the BAROMED key stores the value, and the message PRESSURE
SPAN READING STORED appears on the screen. Refer to Section 5 of
this manual for the complete calibration procedure.
Figure 2-5: CALIBRATION MENU SCREEN
CALIBRATION MENU
O2MED
OFFSET O2 CELL A
0138
0127
OFFSET O2 CELL B
0138
0129
BAROMED
OXYGEN OFFSET READINGS STORED
PRESSURE SPAN (60 cm H2O)
0243
0481
EXIT
TRACE
TREND
Rev. A
2-7
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...