NM2B
PMC PROCEDURE (continued)
6.8.9
Bain Circuit Adapter - if applicable
6.8.9.1
Close the APL valve by turning the knob fully clockwise.
6.8.9.2
Insert the O
2
sensor plug into the O
2
sensor inlet on the Bain
Circuit.
6.8.9.3
Attach a test terminal with a cuff inflation bulb (P/N S010159) to
the Breathing Bag port on the Bain Circuit.
6.8.9.4
Attach a cmH
2
O digital pressure meter to the female quick connect
connection on the Bain Circuit.
6.8.9.5
Occlude the expiration port on the Bain Circuit.
6.8.9.6
Apply 50cmH
2
O to the Bain Circuit via test terminal and inflation
bulb.
( )
6.8.9.7
After 30 seconds, what is the pressure on the cmH
2
O digital
pressure meter? (45 to 50 cmH
2
O)
6.8.9.8
Verify that the pressure indicated on the cmH
2
O gauge is within
3 cmH
2
O of the digital pressure meter reading.
6.8.9.9
Open the APL valve by turning the knob fully counter-clockwise.
6.8.9.10
Connect a test hose from the fresh gas outlet to the Expiration
port of the Bain Circuit.
6.8.9.11
Set the O
2
flow to 10 L/min.
( )
6.8.9.12
What is the pressure on the cmH
2
O digital pressure meter? (
≤
3
cmH
2
O)
6.8.9.13
Verify that the pressure indicated on the cmH
2
O gauge is within
3 cmH
2
O of the digital pressure meter reading.
6.8.9.14
Remove the test terminal and inflation bulb from the Breathing
Bag port.
6.8.9.15
Return all controls to their original positions.
Rev. Y
6-21
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...