NM2B
PMC PROCEDURE (continued)
6.24A
BELLOWS: ADULT - if applicable
6.24A.1
Set the FREQUENCY to 10 BPM.
6.24A.2
Set the I:E RATIO to 1:2.
6.24A.3
Adjust the O
2
flow to 300 ml/min.
6.24A.4
Adjust the INSPIRATORY FLOW to MED.
6.24A.5
Adjust the Tidal Volume to 200 ml.
6.24A.6
Press the O
2
Flush momentarily to inflate the bellows.
( )
6.24A.7
What is the Tidal Volume on the test volumeter? ___ ml (125-250)
6.24A.8
Adjust the Tidal Volume to 1000 ml.
6.24A.9
Press the O
2
Flush momentarily to inflate the bellows.
( )
6.24A.10 What is the Tidal Volume on the test volumeter? ___ ml (900-1100)
6.24A.11 Adjust the O
2
flow to 5 l/min.
6.24A.12 Adjust the INSPIRATORY FLOW to maximum.
6.24A.13 Adjust the Tidal Volume to maximum.
( )
6.24A.14 What is the Tidal Volume on the test volumeter? ___ ml (
≥
1400)
6.24B
VENTILATOR RELIEF VALVE: ADULT
6.24B.1
Adjust the O
2
flow to 10 l/min.
6.24B.2
Adjust the INSPIRATORY FLOW to MED.
6.24B.3
Adjust the I:E RATIO to 1:3, and the FREQUENCY to 10.
6.24B.4
Adjust the Tidal Volume to 1200 ml.
( )
6.24B.5
What is the PEEP? ___ cm H
2
O (
≤
3)
6.24B.6
Adjust the O
2
flow to 500 ml.
( )
6.24B.7
Does the ventilator deliver the full Tidal Volume during 5 consecutive
cycles? ___ (Y)
6.24B.8
Does the bellows stop adjust smoothly? ___ (Y)
Rev. Y
6-53
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...