DIAGNOSTICS (continued)
NM2B
Table 2-3: NARKOMED 2B ERROR CODES, GROUP 3
Left
Char.
Error(s)
Right
Char.
Error(s)
0
No Error
0
No Error
1
Backup memory fail (ROM compare test)
1
Clock fail (write/readback test)
2
Backup memory fail (array recall test)
2
Clock fail (original time test)
3
Backup memory fail (ROM compare test)
Backup memory fail (array recall test)
3
Clock fail (write/readback test)
Clock fail (original time test)
4
AC power fail
4
Clock fail (seconds not incrementing)
5
AC power fail
Backup memory fail (ROM compare test)
5
Clock fail (seconds not incrementing)
Clock fail (write/readback test)
6
AC power fail
Backup memory fail (array recall test)
6
Clock fail (seconds not incrementing)
Clock fail (original time test)
7
AC power fail
Backup memory fail (array recall test)
Backup memory fail (ROM compare test)
7
Clock fail (seconds not incrementing)
Clock fail (original time test)
Clock fail (write/readback test)
8
Reserve power fail (batt. low/disconnected)
8
Backup memory fail (checkerboard test)
9
Reserve power fail (batt. low/disconnected)
Backup memory fail (ROM compare test)
9
Backup memory fail (checkerboard test)
Clock fail (write/readback test)
A
Reserve power fail (batt. low/disconnected)
Backup memory fail (array recall test)
A
Backup memory fail (checkerboard test)
Clock fail (original time test)
B
Reserve power fail (batt. low/disconnected)
Backup memory fail (array recall test)
Backup memory fail (ROM compare test)
B
Backup memory fail (checkerboard test)
Clock fail (original time test)
Clock fail (write/readback test)
C
Reserve power fail (batt. low/disconnected)
AC power fail
C
Backup memory fail (checkerboard test)
Clock fail (seconds not incrementing)
D
Reserve power fail (batt. low/disconnected)
AC power fail
Backup memory fail (ROM compare test)
D
Backup memory fail (checkerboard test)
Clock fail (seconds not incrementing)
Clock fail (write/readback test)
E
Reserve power fail (batt. low/disconnected)
AC power fail
Backup memory fail (array recall test)
E
Backup memory fail (checkerboard test)
Clock fail (seconds not incrementing)
Clock fail (original time test)
F
Reserve power fail (batt. low/disconnected)
AC power fail
Backup memory fail (array recall test)
Backup memory fail (ROM compare test)
F
Backup memory fail (checkerboard test)
Clock fail (seconds not incrementing)
Clock fail (original time test)
Clock fail (write/readback test)
2-5A
Rev. L
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...