NM2B
PMC PROCEDURE (continued)
6.15 FLOWMETERS
( )
6.15.1
Oxygen Flowmeter
6.15.1.1
Open the O
2
cylinder valve.
6.15.1.2
Is it possible to adjust the flow of oxygen over the full range of the
flowmeters? ___ (Y)
6.15.1.3
Close the O
2
cylinder valve and bleed the pressure.
6.15.1.4
Connect the O
2
pipeline supply, and verify the operation of the
oxygen flowmeter.
6.15.1.5
Is the correct flow control knob and label attached to the oxygen
flow control valve? ___ (Y)
6.15.1.6
Close the oxygen flow control valve.
*
6.15.1.7
Set the Gas Selector switch to the O
2
+ N
2
O position.
( )
6.15.1.8
What is the minimum flow of oxygen? ___ ml (100-200 ml/min);
(0 ml for 4107615A (Minimum O
2
Flow Elimination))
( )
6.15.2
Nitrous Oxide Flowmeter
6.15.2.1
Set the oxygen flow to 4 l/min.
6.15.2.2
Open the nitrous oxide cylinder valve.
6.15.2.3
Is it possible to adjust the flow of nitrous oxide over the full range
of the flowmeter? ___ (Y)
6.15.2.4
Close the nitrous oxide cylinder valve and bleed the pressure.
6.15.2.5
Connect the N
2
O pipeline supply, and verify the proper operation
of the N
2
O flowmeter.
6.15.2.6
Is the correct flow control knob and label attached to the N
2
O flow
control valve? ___ (Y)
6.15.2.7
Close the oxygen and nitrous oxide flow control valves.
*
These items do not apply to 2-gas machines or later design (P/N 4113329-001)
ORC.
Rev. Y
6-35
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...