PMC PROCEDURE (continued)
NM2B
6.8.5
Breathing System Leak
6.8.5.1
Verify the System Power switch is at STANDBY.
6.8.5.2
Close all flow control valves.
6.8.5.3
Interconnect the inspiratory and expiratory valves with a 12-inch
hose.
6.8.5.4
Attach a test terminal with a cuff inflation bulb (P/N S010159) to
the bag mount.
6.8.5.5
Set the Man/Auto selector valve to BAG.
6.8.5.6
Close the APL valve.
6.8.5.7
Apply 50 cm H
2
O pressure to the absorber system.
( )
6.8.5.8
After 30 seconds, what is the pressure in the absorber system? ___
cm H
2
O (>30)
6.8.6
APL Valve
6.8.6.1
Open the APL valve to its stop.
6.8.6.2
Turn the SYSTEM POWER switch to ON.
6.8.6.3
Set the oxygen flow to 8 l/min.
( )
6.8.6.4
What is the pressure on the absorber pressure gauge? ___ cm H
2
O
(<3)
6.8.6.5
Close the oxygen flow control valve, turn the System Power switch
to STANDBY, and remove the test terminal from the bag mount.
6.8.7
Absorber Flow Direction and Leak
6.8.7.1
Expiration Valve Leak
6.8.7.1.1 Close the APL valve.
6.8.7.1.2 Connect a 22mm hose between the inspiration valve and the
bag mount.
6.8.7.1.3 Connect a test terminal to the expiration valve.
6.8.7.1.4 Connect a Capnomed flowmeter to the test terminal.
Rev. Y
6-18
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...