NM2B
PMC PROCEDURE (continued)
( )
6.11.2
Cylinder Gauges
6.11.2.1
Are the pressure gauges correct for the gases indicated by the
flowmeters? ___ (Y)
6.11.2.2
Is the gauge closest to the table top for cylinder supply pressure?
___ (Y)
6.11.2.3
Bleed all pressure from the cylinder circuit.
6.11.2.4
Are the cylinder gauges at zero (0) PSI? ___ (Y)
6.11.2.5
Open the cylinder valves.
6.11.2.6
Do the cylinder pressure gauges respond properly? ___ (Y)
( )
6.11.3
Pipeline Gauges
6.11.3.1
Are the gauges right below the flowmeters for pipeline supply
pressure? ___ (Y)
6.11.3.2
Are the pipeline pressure gauges at zero (0) PSI? ___ (Y)
6.11.3.3
Connect the pipeline supply.
6.11.3.4
Do the pipeline pressure gauges respond properly? ___ (Y)
6.11.3.5
Are the correct gas identification labels affixed at each of the
pipeline inlets? ___ (Y)
6.11.3.6
Does the back panel identify each of the pipeline inlets properly?
___ (Y)
6.11.3.7
Disconnect the pipeline hoses.