PMC PROCEDURE (continued)
NM2B
6.17 OXYGEN CONCENTRATIONS
6.17.1
Nitrous Oxide Concentration Test
6.17.1.1
Verify the oxygen flow is at 4 l/min.
6.17.1.2
Depress the O
2
Flush button for 5 seconds.
6.17.1.3
Does the O
2
Med read 97-100% after the value stabilizes?__(Y)
6.17.1.4
Set the nitrous oxide flow to 2 l/min.
( )
6.17.1.5
After the value stabilizes, what is the oxygen concentration? ___
% (64-70)
6.17.1.6
Close the nitrous oxide flow control valve.
6.17.2
Air Concentration Test - if applicable
6.17.2.1
Depress the O
2
FLUSH button for 5 seconds.
6.17.2.2
Does the O
2
Med read 97-100% after the value stabilizes?__(Y)
*
6.17.2.3
Set the gas selector switch to ALL GASES.
6.17.2.4
Verify the O
2
flow is at 4 l/min; set the air flow to 2 l/min.
( )
6.17.2.5
After the value stabilizes, what is the O
2
concentration? ___%(71-
77)
6.17.2.6
Close the air cylinder valve and bleed the pressure from the circuit
- if applicable.
6.17.2.7
Close the air flow control valve.
*
6.17.2.8
Set the gas selector switch to O
2
+N
2
O.
* These items do not apply to machines with later design (P/N 4113329-001) ORC.
Rev. Y
6-40
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...