NM2B
PMC PROCEDURE
6.0 PMC PROCEDURE, NARKOMED 2B
The procedures in this section shall be performed in their entirety each time a
component is removed, replaced, calibrated, adjusted and during all scheduled Periodic
Manufacturer’s Certification (PMC) visits. A PMC Checklist form, P/N S010211,
available from the Draeger Medical, Inc. Technical Service Department, shall be
completed by the TSR each time a PMC is performed. Steps in the procedure marked
with ( ) require a response at the corresponding line on the checklist form.
Space is also provided on the PMC checklist form to record the results of a vapor
concentration test.
NOTE:
Test equipment listed below with an asterisk (*) requires calibration at a
maximum interval of one year. Verify the dates on test equipment calibration
labels. DO NOT USE any test equipment having an expired calibration date.
Notify your supervisor immediately if any equipment is found to be out of
calibration.
In the space provided at the bottom of the PMC checklist form, record the
Model and ID number of all calibrated test equipment used. Also record the
calibration due dates.
Test Equipment Required:
*-- Multi-Meter (Fluke or equivalent)
*-- Electrical Safety Analyzer (Biotek 501 Pro or equivalent)
*-- Regulator Test Pressure Gauge, P/N S000063 or equivalent
-- Fresh Gas Outlet Volume Test Device, P/N S010158 or equivalent
-- Fresh Gas Leak Test Device, P/N 4113119 or equivalent
-- Adapter Assembly, Test Terminal, P/N 4104389 or equivalent: two are required
*-- Flowmeter Test Stand (Capnomed), P/N S000081 or equivalent
-- Breathing System Leak Test Device, P/N S010159 or equivalent
-- Baromed Pressure Test Fixture or equivalent
*-- Test Minute Volume Meter, P/N 2212300 or equivalent
*-- Digital Pressure Manometer (SenSym PDM 200CD or Equivalent)
*-- Riken Gas Indicator, Model 18H, or 1802D or equivalent
-- Stop Watch
Materials Required:
-- Dow Corning High Vacuum Grease, P/N S4105908
-- Spiromed Sensor Lubrication Kit, P/N 2218180
-- Breathing Bag, 3 liter, P/N 9995330 or equivalent
-- Tube, Corrugated, 22 mm x 12 in. long, P/N 9995112
Test equipment illustrations are shown on the following pages.
Rev. Y
6-1
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...