PMC PROCEDURE (continued)
NM2B
( )
6.9
MANUAL SPHYGMOMANOMETER - if applicable
6.9.1
Insert the male Luer fitting of the Sphygmomanometer squeeze bulb-
hose assembly into the female Luer fitting labeled BP BULB on the
front of the machine.
6.9.2
Wrap the blood pressure cuff around an "E" cylinder.
6.9.3
Hand-pump the squeeze bulb until pressure of 200 mm Hg is indicated
on the Sphygmomanometer gauge on the machine. Pinch the hose
adjacent to the Luer fitting to assure that the hose-bulb is not the
source of any leak.
6.9.4
After
thirty
(30)
seconds,
what
is
the
pressure
on
the
Sphygmomanometer gauge? ___ mm Hg (190-200)
6.9.5
Bleed the pressure.
6.9.6
The Sphygmomanometer should indicate within the band.
6.9.7
Attach a test gauge in series with the sphygmomanometer gauge being
tested.
6.9.8
Hand-pump the squeeze bulb until the machine gauge indicates 100
mm Hg.
6.9.9
What does the test gauge indicate? ___ mm Hg (90-110).
6.9.10
Hand-pump the squeeze bulb until the machine gauge indicates 200
mm Hg.
6.9.11
What does the test gauge indicate? ___ mm Hg (180-220).
6.9.12
Hand-pump the squeeze bulb until the machine gauge indicates 300
mm Hg.
6.9.13
What does the test gauge indicate? ___ mm Hg (210-390).
6.9.14
Remove the test gauge.
6.9.15
Remove the blood pressure cuff from the "E" cylinder.
Rev. Y
6-22
Содержание Narkomed 2B
Страница 2: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 18: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 32: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 240: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 244: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 306: ...RETURN TO THIS MANUAL S TABLE OF CONTENTS RETURN TO CD ROM TABLE OF CONTENTS ...