PMC PROCEDURE (continued)
NM2B
( )
6.2
SELF-DIAGNOSTICS
6.2.1
Connect the pipeline supply or open the cylinders.
6.2.2
Turn the System Power switch to ON.
6.2.3
Verify that the following is displayed on the alarm CRT:
VIDEO TEST
PASS
NARKOMED 2B
FIRMWARE TEST
PASS
VERSION x.xx DIAGNOSTICS
MEMORY TEST
PASS
COPYRIGHT, NAD INC. 1987-94
TIMERS TEST
PASS
ANALOG TEST
PASS
AUDIO TEST - PRIMARY
PASS
- BACKUP
PASS
SERIAL I/O TEST
PASS
CLOCK TEST
PASS
BACKUP MEMORY TEST
PASS
AC POWER TEST
PASS
RESERVE POWER TEST
PASS
FUNCTIONAL
( )
6.2.4
Record the machine software version on the header of the checklist
form.
Rev. Y
6-4