TROUBLESHOOTING GUIDE (continued)
NM2B
GUIDE 1: Loss of O
2
Monitor
N
Y
N
Y
N
Y
N
Y
Y
N
N
Y
START
UNIT IS
FUNCTIONAL
DOES UNIT
PASS PMS?
PERFORM A
COMPLETE PMS
ON UNIT
CONTACT NAD
SERVICE DEPT.
REINSTALL ORIGINAL
PROCESSOR ASSEMBLY
IS
OXYGEN ANALYZER
FUNCTIONAL?
REINSTALL ORIGINAL
OXYGEN SENSOR
------------------------
REPLACE
PROCESSOR ASSEMBLY
AS OUTLINED IN
PROCEDURE 4.26
IS
OXYGEN ANALYZER
FUNCTIONAL?
CONNECT CABLE TO
J10 ON PROCESSOR
BOARD (ALSO
CHECK CONTINUITY OF
INTERNAL CABLE)
CONNECT OXYGEN
SENSOR CABLE TO
INTERFACE PANEL
REPLACE
OXYGEN SENSOR
ASOUTLINED IN
PROCEDURE 4.30
IS CABLE
CONNECTED TO
J10 ON PROCESSOR
BOARD?
IS
"O2 SENS DISC"
DISPLAYED ON
CENTRALERT ALARM
DISPLAY?
IS
OXYGEN ANALYZER
FUNCTIONAL?
Rev. A
3-4