Interface Overview
68
©2020 CAE 905K710052 v.1.1
CAE
Air1
When all tests are successfully completed the Self Test is shown as Passed.
Self Test Passed
Tap the
Close
button to return to the Home screen.
When any of the sub-tests do not successfully pass, the Self Test is shown as Failed.
Self Test Failed
IMPORTANT:
While the Self Test can be repeated, if the ventilator continues to fail, it cannot be used
and must be taken out of service. In the case that malfunctions are responsible for the test failures,
the Fail Safe screen will be displayed when the Test Failed screen is closed. Refer to the
Troubleshooting
section for more information.