Appendix B - Alarms
132
©2020 CAE 905K710052 v.1.1
CAE
Air1
Vitals Alarm Criteria and Default Threshold Values
For Vitals alarm conditions, the priorities are defined by the duration of the alerting state as read by
the vitals sensors. For an alert duration of three or more respirations (or 15 seconds), a High priority
alarm is indicated.
Note:
For alarms that are directly related to a control parameter, when the control parameter is
adjusted, the alarm thresholds are temporarily expanded to include the range between the
parameter’s current value and its new set value.
Alarm
Criteria
Default Value per Mode
Apnea
> alarm threshold
All modes: 20s
FiO
2
High
> alarm high threshold for ≥ 3 breaths
All modes: +5% of set FiO
2
value
FiO
2
Low
< alarm low threshold for ≥ 3 breaths
All modes: -5% of set FiO
2
value
Ppeak High
> alarm high threshold for 50 ms
PC mode: +10 cmH
2
O of set Pi value
PS mode: +10 cmH
2
O of set Psupp value
VC mode: 40 cmH
2
O
Ppeak Low
< alarm low threshold for ≥ 3 breaths
PC mode: -10 cmH
2
O of set Pi value
PS mode: -10 cmH
2
O of set Psupp value
VC mode: 7 cmH
2
O
MVe High
> alarm high threshold for ≥ 3 breaths
PC and PS mode: 40 L/min
VC mode: +15% of set value of Vt x RR)
MVe Low
< alarm low threshold for ≥ 3 breaths
PC and PS mode: 5 L/min
VC mode: -15% of set value of Vt x RR)
RR Total
High
> alarm high threshold for ≥ 3 breaths
All modes: +10 of set RR value
Note: Applies to RR backup for PS mode
RR Total
Low
< alarm low threshold for ≥ 3 breaths
All modes: -10 of set RR value
Note: Applies to RR backup for PS mode
PEEP High
> alarm high threshold for ≥ 3 breaths
All modes: +5 cmH
2
O of set PEEP value
PEEP Low
< alarm low threshold for ≥ 3 breaths
All modes: -5 cmH
2
O of set PEEP value
VTe High
> alarm high threshold for ≥ 3 breaths
PC and PS mode: 2000 ml
VC mode: +15% of set Vt value
VTe Low
< alarm low threshold for ≥ 3 breaths
PC and PS mode: 400 ml
VC mode: -15% of set Vt value
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...