Care and Maintenance
112
©2020 CAE 905K710052 v.1.1
CAE
Air1
Preventive Maintenance Schedule
Operators may perform the following preventive maintenance tasks. However, only qualified service
personnel are to perform system maintenance and those preventive maintenance tasks indicated in
the Service Personnel Preventive Maintenance Schedule table.
Operator Preventive Maintenance Schedule
Task
Part
Interval
Clean and disinfect.
Ventilator Exterior Surface
As required
Replace patient circuit and
expiratory filter.
Calibrate the flow sensor after
replacing expiratory filter.
Breathing Circuit and
Inspiratory/Expiratory Filters
Between patients or
according to local protocol
Run preoperational self-
checks.
Entire Ventilator
Between patients or
according to local protocol
Empty any water from
breathing circuit.
Inspect parts for damage.
Replace as necessary.
Breathing Circuit
Every 2 days or according to
local protocol
Visually check or replace the air
inlet filter i
f necessary.
Air Inlet Filter
Monthly. If operating in a
dusty environment, replace
more frequently to prevent
clogging
Visually inspect the cooling fan
filter for dust build-up. If dirty,
replace the air filter.
Cooling Fan Filter
Monthly
Visually inspect the exhaust air
filter for dust build-up. If dirty,
use a cloth tissue to remove
the dust build-up on the filter
mesh.
Exhaust Air Filter
Monthly
Replace air inlet filter
.
Air Inlet Filter
Every 6 months
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...