©2020 CAE 905K710052 v.1.1
143
Appendix D - Ventilator Fault Logs
Alarm Log
The following table identifies each of the codes shown in the alarm log, including:
• Description
• Alarm Level, and
• Operator Action
Note:
Alarm codes begin with 1000 and end at 2701, however, codes do not exist for all sequential
numbers.
ID
Description
Alarm Level Operator Action
1000
Ventilator
Malfunction
High
Hardware Malfunction with ID 1000 to 1099 raised.
Verify Platform log for user action.
1100
Ventilator
Malfunction
High
Hardware Malfunction with ID 1100 to 1199 raised.
Verify Platform log for user action.
1200
Ventilator
Malfunction
High
Hardware Malfunction with ID 1200 to 1299 raised.
Verify Platform log for user action.
2000
Software
Malfunction
High
Shutdown and restart the unit. If the alarm persists
contact CAE customer service.
2001
Gas Supply
Failure
High
Verify Oxygen tank. Requires attention of medical
personnel.
2002
Touch
Screen
Failure
High
Shutdown and restart the unit. If the alarm persists
contact CAE customer service.
2010
Battery
Power
Low
UPS disconnected from the wall. Connect UPS to
the wall. Requires attention of medical personnel.
2011
Battery Low
Medium
UPS disconnected from the wall. Connect UPS to
the wall. Requires attention of medical personnel.
2012
Battery
Critical
High
UPS disconnected from the wall. Connect UPS to
the wall. Requires attention of medical personnel.
2030
Breathing
Circuit
Disconnect
High
Verify if breathing tube is properly placed. Requires
attention of medical personnel.
2031
Breathing
Circuit
Obstruction
High
Verify if breathing tube is properly placed. Requires
attention of medical personnel.
2091
VTe High
High
Requires attention of medical personnel.
2092
VTe Low
High
Requires attention of medical personnel.
2100
Apnea
High
Requires attention of medical personnel.
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...