Troubleshooting
Fail Safe and Pressure Safe Modes
The automated ventilator responses to a mechanical failure or an over-pressurization event are
described in the following table.
Touchscreen is black or fro
-
zen.
• Ventilator was powered
off/on too quickly
• Power off the ventilator for a
minimum of 10 seconds,
then power on again.
Fail Safe displayed at Power
On.
• Ventilator was powered
off/on too quickly
• Power off the ventilator for a
minimum of 10 seconds,
then power on again.
Safety Mode
Triggers
Ventilator Responses
Fail Safe
• Mechanical failure wherein the safe
operating parameters of the
ventilator cannot be maintained.
• Repeated power surges from MCU.
• Ventilation halted and valves
open to create air pathway and
allow patient to breathe room air.
• High priority alarm indications.
• Fail Safe screen displayed on
touchscreen.
Pressure Safe
• Ppeak high alarm persists for three
respiratory cycles or 15 seconds.
• Plimit is reached ≥ 200 ms.
• Pressure safety valve opens to
relieve pressure.
• High priority alarm indications.
• Pressure Safe mode persists for
three seconds and then resumes
ventilation with the previous
settings.
• Pressure Safe mode may occur an
unlimited number of times.
Symptom
Probable Cause
Remedy
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...