Electromagnetic Compatibility
This section provides guidance for Electromagnetic Compatibility (EMC) for basic safety.
The emission characteristics of this ventilator make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is
normally required) this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as relocating or re-
orienting the equipment.
Electromagnetic Emission
Electromagnetic Immunity
Emission Test
Compliance Level
Environment and Guidance
Radiated RF Emissions
CISPR 11
Class A
Group 1
The ventilator suitable for use in industrial
areas and hospitals.
The ventilator uses RF energy only for its
internal functions. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Harmonic Emissions
IEC 61000-3-2
Not Applicable
Voltage Change &
Fluctuations
Flicker Emissions
IEC 61000-3-3
Not Applicable
Emission Test
Compliance Level
Environment and Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±8 kV contact discharge
±8,10, 12 kV α air
discharge
SPECIAL ENVIRONMENT: The relative
humidity should be at least 30% if floors are
covered with synthetic material.
Immunity Proximity
Fields RF Wireless
Communications
Equipment
IEC 61000-4-3
See Table 1
The frequencies and services listed in Table
1 are representative examples that are
based on RF communications equipment in
use at the time of publication of this
collateral standard. The test specification
does not attempt to cover every frequency
and service used in every country.
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...