Appendix A - Parameter Descriptions
128
©2020 CAE 905K710052 v.1.1
CAE
Air1
FiO
2
(Fraction of Inspired O
2
)
The FiO2 parameter controls the concentration of oxygen in the air inspired by the patient.
Default
: 21%
Range
: 21% - 100%
I:E ratio (Inspiratory Time) / I:E Backup
The Inspiratory Time parameter control is used to increase or decrease the inspiratory time and
impacts the inspiratory time:expiratory time ratio (I:E ratio).
Range
:
1:1 - 1:4
Resolution
:
0.1
RR (Respiratory Rate) / RR Backup
The RR parameter is used to set the respiratory rate of the patient to a given number of breaths per
minute and impacts the inspiratory time:expiratory time ratio (I:E ratio).
Default
: 15 breaths per minute
Range
: 4 - 50 breaths per minute
PSafety (Maximum Limited Pressure)
The ventilator includes protection to prevent the airway pressure from exceeding a maximum
specified limited pressure (Psafety) under both normal conditions and a single fault condition. In
Volume Control mode, the Psafety shall be set directly. In Pressure Control and Pressure Support
modes, the PSafety shall be 10 cmH
2
O above the high pressure limit.
Default
: 50 cmH
2
O
Range
: Up to 80 cmH
2
O
Tapnea (Apnea Time)
When in Pressure Support mode, Tapnea is set. When Tapnea is reached, the ventilator will fall back
into a control breath mode at the set backup values of Pi, RR and I:E.
Default
: 20 seconds
Range
: 15 - 45 seconds
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...