©2020 CAE 905K710052 v.1.1
137
Appendix C - Operation Modes
A
PPENDIX
C - O
PERATION
M
ODES
The CAE Air1 ventilator supports three distinct modes of operation. These modes define the
breathing pressure, volume and degree of breathing assistance that a patient receives from the
ventilator. The modes are selectable through the graphical user interface on the touchscreen.
For all three modes, the ventilator ensures that the airway pressure never drops below the set PEEP
value. The concentration of oxygen in the air delivered to the patient is determined by the set FiO
2
value (when a pressurized oxygen source is connected to the ventilator).
The following table describes the three modes and identifies the parameters that can be set for each.
Note:
Pi is defined as the inspiratory pressure relative to atmospheric pressure
Mode
Description
Parameters
Pressure
Control
(PC)
The ventilator delivers a breath to the patient based
on the set Respiratory Rate and the detection of an
inspiratory effort from the patient above the
pressure trigger threshold.
The ventilator increases the pressure delivered until
the set Pi value is met. The flow pattern decelerates
to maintain the Pi.
The pressure is maintained for a duration
determined by the set RR and I:E ratios.
• Pi
• PEEP
• RR
• I:E
• Pressure Trigger
• FiO
2
Volume
Control
(VC)
The ventilator delivers a breath to the patient based
on the set Respiratory Rate and the detection of an
inspiratory effort from the patient above the
pressure trigger threshold.
The ventilator delivers a constant flow until the set
Tidal Volume is achieved.
As a safety mechanism the delivery of flow stops if
the Inspiratory Pressure (Pi) reaches the set Psafety
value.
• Vt
• PEEP
• RR
• I:E
• Pressure Trigger
• FiO
2
• Psafety
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...