Setup
48
©2020 CAE 905K710052 v.1.1
CAE
Air1
If failure recurs, tap
Settings
to navigate to the Settings screen to view the test failure logs.
This action must be performed by a trained technician with administrator privileges. Refer
to the
Care and Maintenance
section for more information.
Home Screen with Tests Failed Notification
Step 8: Perform O
2
Calibration
If pressurized oxygen is being used (see
Step 2
), the O
2
Calibration must be performed prior to the
first use of CAE Air1.
To perform O
2
Calibration:
a. Make sure the ventilator is connected to a pressurized O
2
source.
b. Make sure there is no patient connected.
c. Tap the
Settings
button on the Home screen.
d. Select
Start the O
2
Calibration
on the Settings screen.
Calibration will automatically be performed.
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...