©2020 CAE 905K710052 v.1.1
129
Appendix A - Parameter Descriptions
Tidal Volume
The Tidal Volume setting is for the volume of gas intended to enter the lung during a respiratory
cycle.
Default
: 500 mL
Range
: 200 - 1500 mL
Accuracy Tolerances
All tolerances are checked against the ASL 5000 in BTPS unit.
Volume Control Accuracy
VTe:
±30 ml
PEEP:
±0.9 cmH
2
O
FiO
2
:
±((7.2% of the set FiO
2
) – 0.15) %FiO2 for 300
VT
500mL,
and ±7.5% of the set FiO
2
for VT < 300mL
Pressure Control Accuracy
Ppeak:
±2.6 cmH
2
O
PEEP:
±(1.1 -(4% of the set PEEP)) cmH
2
O
FiO
2
:
±(1.9 +(6% of the set FiO
2
)) %
Oxygen Concentration Rate
This section describes the response of the ventilator to an increase in the set oxygen concentration.
The length of time required for the oxygen concentration in the tidal volume range to change from a
volume fraction of oxygen of 21% to 90% is defined below:
Vtidal ≥ 300 ml ≥ 82 sec (1 minute 22 seconds)
300 ml ≥ Vtidal ≥ 200 ml ≥ 153 sec (2 minutes 33 seconds)
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...