©2020 CAE 905K710052 v.1.1
61
Interface Overview
There are five tests included in the Self Test, each with its own sequence of screens:
1. Primary Alarm System
2. Backup Alarm System
3. Power Source Switch
4. Component Functional Check
5. Software Integrity Check
For each of the five tests, follow the onscreen instructions.
As each test is passed, a green checkmark appears beside the test name on the left side of the
screen. If a test fails, a red X will appear beside the test name instead.
The Self Test is determined to Pass only if all of these sub-tests are passed.
Primary Alarm System
The Primary Alarm System test verifies the functioning of the ventilator’s audible and visible alarms.
Follow the onscreen instructions:
Primary Alarm System Test - Start Screen
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...