©2020 CAE 905K710052 v.1.1
127
Appendix A - Parameter Descriptions
A
PPENDIX
A - P
ARAMETER
D
ESCRIPTIONS
This section describes the parameter ranges for the Control widgets and accuracy tolerances used in
the CAE Air1 ventilator.
Control Widgets
Parameter values can be adjusted and set to meet the patient’s changing ventilation needs.
Pi (Inspiratory Pressure) / Pi Backup
The Pi parameter controls the inspiratory pressure of a mandatory breath applied by mechanical
ventilation.
Default
: 15 cmH
2
O
Range
: 10 - 50 cmH
2
O
PSupp (Support Pressure)
The Psupp parameter controls the support pressure for a spontaneous patient breath.
Default
: 15 cmH
2
O
Range
: 10 - 50 cmH
2
O
PEEP (Positive End Expiratory Pressure)
The PEEP parameter controls the positive end-expiratory pressure applied during mechanical
ventilation to prevent alveolar collapse and improve oxygenation.
Default
: 5 cmH
2
O
Range:
0 - 30 cmH
2
O
Pressure Trigger
The pressure trigger parameter controls the sensitivity of the ventilator to detect a patient's
spontaneous breathing effort.
Default
: 2.0 cmH
2
O
Range
: 0.5 - 20 cmH
2
O
Содержание Air1
Страница 1: ...CAEAir1 Mechanical Ventilator User Guide...
Страница 7: ...vi 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 9: ...Specifications 2 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 33: ...Safety 26 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 37: ...Introduction 30 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 45: ...Equipment Overview 38 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 80: ...2020 CAE 905K710052 v 1 1 75 Interface Overview Setup Screen PC Mode Setup Screen PS Mode Setup Screen VC Mode...
Страница 131: ...Troubleshooting 126 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 135: ...Appendix A Parameter Descriptions 130 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...
Страница 153: ...Appendix F Applicable Standards 148 2020 CAE 905K710052 v 1 1 CAEAir1 INTENTIONALLY LEFT BLANK...