Calibration | 5-7
When performing a primary calibration or the first calibration after a
primary calibration, the reading and the setpoint values often won’t
match. Once the first regular calibration after a primary calibration is
performed, the readings should match.
You are then prompted to calibrate the zero gases:
O2 xx.x % Set xx.x %
Press ENTER to Zero
5. Press
Enter
when the zero gas reading(s) on the left of the display stabi-
lizes. The reading on the far right of the first display line shows the setpoint
value you entered for the O
2
zero gas using the
Cal Gas Values
menu
option. This O
2
zero gas setpoint value should match the cal gas cylinder
for the O
2
zero gas connected to your RCU.
Then, if you specified a recovery time, the system will display the recov-
ery time remaining:
Recovery xx:xx
System Calibrating
The system returns to monitoring process readings.
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...