A-12 | Thermox Series 2000 WDG Insitu Analyzer
05/05
Line1_flg
What is displayed on line 1 of the display
0 – blank line
2 – oxygen
4 – cell temperature
6 – cell mv
8 – thermocouple mv
10 – box temp
IQ
12 – combustibles
14 – methane
16 – methane and combustibles
18 – combustible voltages
IQ
20 – methane voltages
IQ
22 – time and date
24 – user text
S2
AF Ratio
IQ
26 – WOBBE
IQ
R/W
06/06
Line2_flg
What is displayed on line 2 of the display
0 – blank line
2 – oxygen
4 – cell temperature
6 – cell mv
8 – thermocouple mv
10 – box temp
IQ
12 – combustibles
14 – methane
16 – methane and combustibles
18 – combustible voltages
IQ
20 – methane voltages
IQ
22 – time and date
24 - user text
S2
AF Ratio
IQ
26 - WOBBE
IQ
R/W
07/07
Line3_flg
What is displayed on line 3 of the display
0 – blank line
2 – oxygen
4 – cell temperature
6 – cell mv
8 – thermocouple mv
10 – box temp
IQ
12 – combustibles
14 – methane
16 – methane and combustibles
18 – combustible voltages
IQ
20 – methane voltages
IQ
22 – time and date
24 - user text
S2
AF Ratio
IQ
26 – WOBBE
IQ
R/W
08/08
Oxygen
The oxygen reading in percent
Read only
09/09
Cmbcomb
The combustibles reading
Read only
0A/10
Mthcomb
The methane reading
Read only
0B/11
Cell_temp
The current cell temperature in degrees C
Read only
0C/12
Cell_mv
The cell millivolts
Read only
0D/13
Tc_mv
The thermocouple millivolts
Read only
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...