Controller / User Interface | 4-13
Analog Range Key
Set Current
Range
Set Track/Hold
Set Current
Mode
Output Filtering
Select Function
ANALOG RANGE KEY
Figure 4-3. Analog Range Key menu.
You can direct process readings to either of the two analog output ports
on the rear of the control unit. If you have the combustibles option, you
can also have process readings directed to two additional analog output
ports activated with this option. All functions in this chapter are directed
to a specific analog output port.
ANALOG KEY
FUNCTIONS/OPTIONS
Set Current Range
Define what information should be sent to the analog
output port during a calibration. For example:
4% O
2
= 4 mA
10% O
2
= 20 mA
Set Track/Hold
Hold output at last process reading during a cal or
verify,
or track the calibration readings.
Set Current Mode
Define analog output range:
0 to 20 mA
4 to 20 mA
Output Filtering
Smooth out variations in readings
Select Function
Define the information the analog output should track:
% O
2
Cell temperature
Cell Millivolts
Thermocouple Millivolts
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...