6-12 | Thermox Series 2000 WDG Insitu Analyzer
General Troubleshooting
Your system may pass calibrations, yet still seem to be reading incorrect
oxygen levels. If this is the case, you may want to check the following:
Leak Check
This check does not apply to Insitu users.
Leaks can lead to inaccurate readings, especially if operating under a sig-
nificant pressure or vacuum.
• Check that all compression fitting and pipe thread connections are
leak tight.
• Be sure that the mounting plate or mounting flange gasket on the rear
of the sensor is in place (see the “Sensor Mounting” section in Chapter
2 for details on the placement of the gasket).
Sniffing for Leaks
For processes under vacuum you can check for leaks by sniffing the fit-
tings with another gas (for example, nitrogen or pure oxygen), being sure
to avoid the area over the top of the cell.
1. Use a piece of tygon or plastic tubing with a 1/8” stainless steel nozzle
to apply the gas from a cylinder (using stainless steel prevents any
problems that might occur with plastic melting the nozzle on hot sen-
sor components).
2. Monitor the response from the cell. When the cell millivolt reading
changes, it indicates a leak in that area of the plumbing (the vacuum
of the process pulls in the gas).
3. If not convenient to view the control unit display, you can also apply a
volt meter to terminals 1 and 2 on the sensor board (Cell + and Cell -)
to see if the cell millivolts change, indicating a leak.
Pressurizing for Leaks
1. Remove the sensor from the process (after allowing sensor to cool)
and pressurize it with 5 lbs. of air, plugging any exiting ports, the inlet
probe, and exhaust tube holes. Then go over the sensor fittings with a
leak detector fluid.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...