4-10 | Thermox Series 2000 WDG Insitu Analyzer
Setting up a primary calibration
Make sure you have properly entered calibration gas values using the
Cal
Gas Values
option from the
Calibrate
key before starting your primary
calibration.
1. From the
Setup
key, select
Primary Calib
.
2. If you have the auto calibration card option (and a remote calibra-
tion unit), you will be prompted to select
Auto Calibrate
,
Remote
Calibrate
, or
Manual Calibrate
. If you don’t have the auto calibration
option, the system defaults to a manual calibration, and you will be
prompted to inject span gas into the calibration gas inlet port.
From this point, operations continue as they do for a regular calibration.
Please see the “Initiate Cal/Initiate Verify” section in Chapter 5 for help
on how to perform the calibration.
If you have the combustibles option, during a primary calibration the
system will not use the combustible span gas to calibrate the system.
When the primary calibration is finished, the system will notify you
that a standard calibration is required by displaying a
Calibration
Required
message.
Oxygen readings on the display will not be accurate until the regular
calibration is performed.
A 15-minute recovery time follows the primary calibration to allow
the cell to stabilize. Perform the regular calibration after this 15 minute
stabilization period,
During the recovery time immediately following the primary calibra-
tion, you may see some temperature related error messages. You can
ignore these messages and assume no problem exists. This is because
the cell temperature is adjusted after a primary calibration.
NOTE
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...