5-18 | Thermox Series 2000 WDG Insitu Analyzer
Define a recovery time so the control unit can return to reading process
gases after reading calibration gases without affecting the analog outputs
or the triggering of alarms.
This menu option also can be used to prevent triggering of alarms if, for
example, your low alarm is set to activate if readings fall below 3%, and
your calibration gas is 2%.
This menu option applies to all automatic, remote, and manual cali-
brations and verifications.
1. From the
Set Cal Timers
menu, select
Recovery Duration
.
You will be prompted to enter a calibration recovery duration:
Cal recovery = xx:yy
Enter new time: Mn:Sc
2. Enter the recovery time, then press
Enter
. If you want to use the cur-
rently displayed recovery time, press
Enter
without entering a new
value.
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...