Controller / User Interface | 4-5
Display
Define what information should appear on the top three (3) display lines
of the control unit. Choices include:
Percent oxygen (% O
2
)
Cell temperature
Cell millivolts
Thermocouple millivolts (T/C millivolts)
Current date/time
User text
Turn off (leave display line blank)
When you access menus, the menu options can occupy display lines 3
and 4, and sometimes display line 2. Therefore, place the information
that is most critical for you to view on the topmost display lines.
Setting up the dIsplay
1. From the
Setup
menu, select
Display
.
2. Select
Display Line #
1,
Display Line #
2, or
Display Line #
3, de-
pending on where you want to place the information.
3. Select the type of information you want to place on the display line.
Entering User Text
The
Down
arrow found on the third line of the display allows you to see
your current position when entering user text. When you first select the
User Text
option, this
Down
arrow moves to above the first position.
To enter text, use the
Up
and
Down
arrow keys on the keypad to scroll
through the ASCII text characters. When the letter you want to enter ap-
pears in position, press the
Right
arrow key to move to the next position.
You can also press the
Left
arrow key to move back one position in your
text message. When you are finished entering your message, press
Enter
.
User text can be up to 20 characters long.
NOTE
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...