4-16 | Thermox Series 2000 WDG Insitu Analyzer
Set Track/Hold Option
Define the signals to send to the selected analog output port during a cali-
bration or verification. You have the option to either hold the last process
reading or track calibration readings. Separate decisions can be made for
verify and calibrate operations, so you can hold the last process readings
during verifications, but track calibration gas readings during calibrations.
Setting track or hold
1. Select an analog output port.
2. From the
Analog Range
key, select
Set Track/Hold.
You can choose to either
hold
process readings or
track
calibration
readings on the analog output port. Separate choices can be made for
calibration and verify operations.
For
calibration
operations, select one of the following:
Hold during cal
or
Track during cal
Your choice will be confirmed on the display:
Out # 1 will TRACK
during Calibration
or
Out # 1 will HOLD
during Calibration
For
verify
operations, select one of the following:
Hold during Verify
or
Track during Verify
Your choice will be confirmed on the display:
Out # 1 will TRACK
during Verify
or
Out # 1 will HOLD
during Verify
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...