A-18 | Thermox Series 2000 WDG Insitu Analyzer
5F/95
Gas_ctr
Which gas is currently flowing to the
sensor
0 – span gas flowing
1 – zero gas flowing
2 – combustible span gas
3 – methane span gas (if 3 span value RCU
installed)
IQ
- Normal operation
S2
4 – Step func. Process gas flowing (1
st
time)
5 – Step func. Span gas flowing
6 – Step func. Zero gas flowing
7 – Step func. Process gas flowing (2
nd
time)
0x80 – User Aborted Cal (set from abort to
recovery completed)
0x81 – Recovery Period (if in a cal or
verify) per loc 60/95.
– Normal operation (if not in a
cal or verify) per loc 60/95.
Read only
60/96
Calflag
Which type of calibration the unit is in:
0 – In normal calibration
1 – In verify
2 – In primary cal
3 – In none of the above, normal operation
Read only
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...