Calibration | 5-5
Initiate Cal and Initiate Verify
Perform an automatic, remote, or manual calibration or verification. These
menu options function almost identically. See the introduction to this
chapter for more information on the difference between a calibration and
a verification.
Automatic calibrations and verifications can also be set to run at prede-
termined frequencies using the cycle times defined in the
Set Cal Timers/
Set Auto Timers
menu option. Once you set these cycle times, it won’t be
necessary to select the
Initiate Cal
or
Initiate Verify
menu options for an
automatic calibration.
The
Auto Calibrate/Auto Verify
and
Remote Calibrate/Remote Verify
se-
lections are only available if you have installed an Auto Calibrate Card in
your Series 2000 control unit and have a Remote Calibration Unit (RCU).
If the Series 2000 control unit detects that these options are not installed,
it will automatically prompt you with a manual calibration or verification
operation when you select either the
Initiate Cal
or
Initiate Verify
menu
options.
This chapter section is divided into discussions on automatic, remote, and
manual calibrations or verifications.
Auto Calibrate/Auto Verify
Before you can start an auto calibration or verification:
1.
Enter the calibration gas values to match the span and zero calibration
gas cylinders you will use to calibrate or verify your system.
2.
Define how long you want each calibration gas to flow through the sen
-
sor before switching the gases. Remember that with an auto calibration
or verification, the RCU automatically switches the span and zero cali
-
bration gases based on the calibration gas duration times you specify.
Be sure to allow enough time for the gases to stabilize when you define
these times. You may want to use the “Inject Cal Gases” option from
the
Calibrate
menu to determine how long it takes for gases from the
remote calibration unit (RCU) to reach the sensor and stabilize and then
use these times.
3.
Define a time period for the system to recover from reading calibration
gas values to reading process gases.
You can start the auto calibrate cycle as follows:
1. Depending on whether you want to perform a calibration or verifica-
tion, select either the
Initiate Cal
or
Initiate Verify
option from the
Calibrate
key menu.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...