5-8 | Thermox Series 2000 WDG Insitu Analyzer
Manual Calibrate/Manual Verify
To perform a manual calibration or verification, follow these steps:
1. Enter the calibration gas values to match the span and zero calibration
gases you intend to use to calibrate or verify your system.
2. Select
Initiate Cal
or
Initiate Verify
from the
Calibrate
key menu, de
-
pending on whether you want to perform a calibration or verification.
3. Select
Manual Calibrate
or
Manual Verify
.
You are prompted to apply the O
2
span gas to the calibration gas inlet port
on the sensor (tubing must be free of oil and dirt):
Inject Span Gas
ENTER to Continue
CANCEL to Abort
4. Press
Enter
when you have applied the calibration gas.
O2 xx.x % Set xx.x %
Press ENTER to Span
5. Press
Enter
when the O
2
reading on the far left of the first display line
stabilizes. The reading on the far right of the first display line shows the
setpoint value you entered using the
Cal Gas Values
menu option. This
setpoint value should match the cal gas cylinder for the O
2
span gas.
If your O
2
reading on the far left of the display has stabilized, yet is not near the setpoint
value on the right of the display, you should check the cal gas value you entered and
ensure that this value matches the cal gas cylinder for the gas you applied to the cal gas
inlet port of the sensor.
You are then prompted to apply the O
2
zero gas to the calibration inlet
port of the sensor:
Inject Zero Gas
ENTER to Continue
CANCEL to Abort
6. Press
Enter
when you have applied this cal gas.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...