Maintenance and Troubleshooting | 6-13
2. If you see bubbles, it indicates a leak. If using this method be sure to
prevent the liquid from reaching the furnace. If the furnace does get
wet, allow it sufficient time to thoroughly dry.
The WDG-IV style sensor has intentional leaks in the aspirator air
inlet fitting and the screw in the back of the aspirator block. If you are
having problems and your process is under positive pressure, these
intentional leaks will need to be sealed.
Plugged Plumbing Check
Exercise care when working on the sensor. Allow the unit to cool and
wear gloves.
This check does not apply to Insitu users.
1. Examine the inlet and exhaust for plugging problems. When possible,
rod out the probe and exhaust.
2. If this doesn’t solve the problem, disassemble the analyzer to locate
the plug. Clean the plumbing using hot water and a bottle brush.
3. When assembly is complete, check for leaks using a leak detecting
liquid.
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...