Controller / User Interface | 4-11
Sensor Config
Select the sensor you will be using with the Series 2000 controller.
When selecting a sensor, use Table 4-3 for the temperature each type of
sensor uses. If your type of sensor is not listed, find the operating tem-
perature for your sensor, then choose the sensor that uses that operating
temperature.
Use caution when selecting the type of sensor used with the Series
2000 Control Unit. Selecting the wrong sensor will damage sensor
components and could invalidate your warranty coverage.
SENSOR
OPERATING TEMPERATURE °C
Insitu Probe
Normal Cell Range
High Cell mV Range
615
650
WDG IV or HP Series
Normal Cell Range
High Cell mV Range
695
824
CEM/O2
Normal Cell Range
High Cell mV Range
695
824
Table 4-3. Sensor temperatures by analyzer type.
Configurating the sensor
1. From the
Setup
menu, select
Sensor Config.
2. Select the sensor you will use with the Series 2000 control unit. If your
sensor is not listed in Table 4-3, call the factory for the correct operat-
ing temperature for your sensor type.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...