Serial Communications
|
A-17
5D/93
Alr1_flg
What process alarm is associated with
0 – oxygen
S2
oxygen, combustibles, or methane
IQ
1 – calibration
2 – verify
3 – in cal/verify
R/W
5E/94
Alm_config
IQ
Configuration for oxygen alarms, high or
low. If a bit is 1, that is the condition
that the indicated alarm will trip on.
bit 0
– high oxygen, process alarm
IQ
bit 1
– low oxygen, process alarm
IQ
bit 2
– high combustible, process alarm
IQ
bit 3
– high methane, process alarm
IQ
bit 4
– high oxygen, alarm #2
IQ
bit 5
– low oxygen, alarm #2
IQ
bit 6
– high combustible, alarm #2
IQ
bit 7
– high methane, alarm #2
IQ
bit 8
– high oxygen, alarm #3
IQ
bit 9
– low oxygen, alarm #3
IQ
bit 10
– high combustible, alarm #3
IQ
bit 11
– high methane, alarm #3
IQ
bit 12
– Future use
bit 13
–Future use
bit 14
– Future use
bit 15
– Future use
R/W
Alm_config
S2
Configuration for oxygen alarms, high or
low. If a bit is 1, that is the condition
that the indicated alarm will trip on.
bit 0
– Future use
bit 1
– Future use
bit 2
– Future use
bit 3
– Future use
bit 4
– Future use
bit 5
–Future use
bit 6
– Oxygen Alarm #3. 0=Low, 1=High
bit 7
– Oxygen Alarm #4. 0=Low, 1=High
bit 8
– Future use
bit 9
– Future use
bit 10
– Future use
bit 11
– Future use
bit 12
– Future use
bit 13
– Future use
bit 14
– Future use
bit 15
– Future use
R/W
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...