3-18 | Thermox Series 2000 WDG Insitu Analyzer
sensor Connections
Make sure that power has been removed from the control unit and
sensor before making sensor connections. See Figure 3-18, Sensor
Board.
This section describes required wiring connections between the control unit and
the sensor. Follow the connection instructions for the applicable options for your
analyzer. All sensor connections require that you access the sensor board (see
Figure 3-18).
Sensor cabling must be twisted pairs in an overall braided shield, or
twisted pairs in rigid metal conduit. Follow all wiring guidelines as
specified in Figure 3-17 in the front of this wiring section.
Connections
to Sensor
Sensor
Board
Connections
to Controller
AC Line Power
Figure 3-18. sensor board.
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...