Specifications | 2-1
SpecificationS
Series 2000 control Unit
Display
Four-line x 20-character vacuum fluorescent. Displays combinations of oxy
-
gen, ppm combustibles, methane, time and date, cell temperature, user-pro
-
grammable text, thermocouple mV or cell mV. Password protection, program
-
mable pressure compensation and context-sensitive help are also provided.
analog output
Two isolated linear current outputs. Select O
2
, cell temperature, thermo
-
couple mV or cell mV. Each output can be 4-20 mA, 0-20 mA, 20-4 mA or
20-0 mA and is fully scalable. Hold or track during calibration and select
degree of damping. Maximum load 1200 Ohms.
alarms
Two independent oxygen alarms, each high or low selectable. One alarm
can be allocated to % oxygen, calibrate or verify. Set relays to energize or
de-energize on alarm.
contact Rating
0.5A, 30V, 10VA max. noninductive load, AC or DC
Diagnostics
Watchdog timer and service alarms. System test for A/D, RAM, EEPROM
and keypad. Display Line 4 reserved for full text error and diagnostic mes
-
sages. Twenty-entry event log.
calibration
Oxygen cell lifetime extender. Calibrate or verify calibration. Store last
calibration and verification data. Selectable calibration gas run time and
process recovery time. Timed automatic calibration with optional remote
calibration unit.
communications
RS-485, 2-way addressable.
environment
Ambient Temp: 14 to 122° F (-10 to 50° C)
Relative Humidity: 10% to 80%, non-condensing
IEC Installation Category II
IEC Pollution Degree 2
Max. Altitude: 2000 meters
power
Requirements
Nominal 115-230 VAC ±10%, 47-63 Hz. max.; 75 VA max.
enclosure
Standard weatherproof NEMA 4 (IP 56) wall/panel mount. Optional GP
(General Purpose) wall mount, GP 19” rack mount, GP panel mount, or
stainless steel weatherproof NEMA 4X (IP 56) wall/panel mount. All are
UL Listed for NEC Class I, Division 2 areas. Purged and explosion-proof
versions also available.
Hazardous area
UL Listed for NEC Class I, Division 2 areas
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...