6-2 | Thermox Series 2000 WDG Insitu Analyzer
System Messages
Message
Explanation
Auto Cal Pending
Indicates that an automatic calibration is pending. Once the current
calibration or verification is complete, the automatic calibration will
begin. This message will only appear if the
Auto Calibration
option
is installed and enabled.
Auto Verify Pending
Indicates that an automatic verification is pending. Once the current
calibration or verification is complete, the automatic verification will
begin. This message will only appear if the
Auto Calibration
option
is installed and enabled.
Calibration Aborted
Indicates that the calibration has been aborted by the user, or the
calibration failed because the span gas or zero gas was out of
range.
Calibration Required
Once you complete a
primary calibration
, this message is displayed
to inform you that a
regular calibration
is required (after a primary
calibration is performed, you must always then perform a regular
calibration). This message will not clear from the display by remov-
ing and restoring power to the control unit. If you see this message,
select the
Calibrate
key and perform a calibration.
Cal Verify Aborted
The verification was successfully aborted by the system at your
request.
Power Down Detected
The system detected that power was removed, such as from a
power outage. Acknowledge this system message by setting or
reading any user-programmable value. This will clear the message
from the display. This message will not clear by removing and restor-
ing power to the unit.
Primary Calibration
Appears when the system is performing a
Primary Calibration
(see
the
Setup
key for further information on primary calibration). The
message continues to be displayed during the recovery period after
the primary calibration.
System Calibrating
Indicates that a system calibration is in progress. It will appear
during the entire automatic calibration cycle or during the recovery
period of a remote or manual calibration.
System Verifying Cal
Indicates that a system verification is in progress. It will appear
during the entire automatic verification cycle or during the recovery
period for a remote or manual verification.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...