Serial Communications
|
A-7
Variable Table
The following pages provide a summary of the variable locations, name,
and descriptions available.
Location
(Hex/Decimal)
Variable Name
Description
Access
00/00
Flags3
Various flags (1=true, 0=false)
bit 0
– sensor is at operating temperature.
bit 1
– Over/under Flag for tuning
temperature
S2
bit 2
– inject cal gas active.
bit 3
– Auto Cal is occurring
S2
bit 4
– Auto Verify is occurring
S2
bit 5
– in a menu
S2
bit 6
– No limit msg.
S2
bit 7
– calibrate or verify is not
permitted.
S2
Read only
01/01
Inf_mess_flag
System and error message flags (1=true, 0 =
false)
bit 0
– No Exception
S2
, Future use
IQ
bit 1
– Future use
bit 2
– Future use
bit 3
– Future use
bit 4
– Power Down Detected
bit 5
– Watchdog Failure
bit 6
– Detector B Fault
IQ
bit 7
– Detector A Fault
IQ
bit 8
– Combustble cal error
bit 9
– Methane cal error
bit 10
– Warming up
bit 11
– RTC Not Running
IQ
bit 12
– Box Under Temp
IQ
bit 13
– Box Over Temp
IQ
bit 14
– Furnace Circuit Open
IQ
bit 15
– Furnace Over Temp
IQ
bit 16
– Temp Rise Failure
bit 17
– Cell is Over Temp
bit 18
– Zero Gas Range Error
bit 19
– Span Gas Range Error
bit 20
– Primary Calibration
bit 21
– Memory is Corrupted
bit 22
– Excessive Cal Error
bit 23
– Calibration Required
bit 24
– Thermocouple Failure
bit 25
– T/C Circuit Failure
bit 26
– Cal Verify Aborted
bit 27
– Calibration Aborted
bit 28
- System Verifying Cal
bit 29
- System Calibrating
bit 30
– Auto Verify Pending
bit 31
– Auto Cal Pending
Read only
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...