Controller / User Interface | 4-17
Set Current Mode
Define either a 0-to-20 or 4-to-20 mA analog output range for the selected
analog output port. This output is dependent on both the range of read-
ings you define and on the function you track on the analog output.
Setting the current mode
1. Select an analog output port.
2. From the
Analog Range
key, select
Set Current Mode.
3. Select
4-to-20 mA Mode
or
0-to-20 mA Mod
e. Default selection is 4-to-
20 mA.
The following will then appear on the display:
Out # y is set for
4 - 20 ma
o
r
Out # y is set for
0 - 20 ma
Output Filtering Option
Filter out quick transitions in readings to allow a smoother analog output.
Setting output filtering
1. Select an analog output port.
2. From the
Analog Range
key, select
Output Filtering.
You will be
prompted to enter an output filtering value:
[100 =FAST 1 = SLOW]
Fltr = XX New #?
where XX is the current filter value.
100 equals no filtering, and 1 equals the highest possible filter value.
The default value is 100 (no filtering).
3. Type a number between 1 and 100, then press
Enter
. Note that you
should enter 01 to enter 1, 07 to enter 7, 70 for 70, etc.
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...