4-24 | Thermox Series 2000 WDG Insitu Analyzer
Watchdog Relay
During normal system operation, the Watchdog Relay operates as
normally closed (system failure causes open contacts).
The Watchdog Alarm relay de-energizes when the system watchdog timer
times out. The watchdog timer will time out and cause the control unit to
reset itself to prevent the control unit from becoming “lost” due to such
things as noise spikes or power surges. You can use this relay to activate a
remote indicator to notify you if the watchdog timer times out.
In addition, the control unit reads the watchdog after reset, and beeps for
a brief time if the reset was caused by a watchdog condition. Pressing the
reset keys simultaneously (hidden keys to the left and right of the
Setup
key) for a long enough time will cause a watchdog. This will indicate as a
beep after the system restarts after releasing the reset keys.
Exception Log
The exception log provides a history of the following types of events:
• System or error messages
• Alarm conditions
• Reset conditions
• Start of calibrations or verifications
Accessing the exception log
•
From the
Alarm
menu, select
Exception Log.
The system displays the most recent exception log entry on line four of
the display. It also displays when the event was recorded on line three.
Line three of the display indicates the order in which events are recorded
into the log, with number 1 being the most recent entry into the log. The
system is able to store the 20 most recent entries in this exception log.
Pressing the
Down
arrow key shows later exception log entries. Pressing
the
Up
arrow key shows more recent exception log entries. Press
Cancel
at any time to exit this menu option.
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...