Definitions
Calibrations vs. Verifications
The Series 2000 control unit software refers to both
calibrations
and
verifi-
cations
. During a verify operation, the calibration gas readings are taken,
and the difference between what the system records for the gas values
and their known values is recorded. Internal calibration parameters are
not changed. Verify operations are often used to check that the instru-
ment is running within required tolerances.
On the other hand, during a calibration software, algorithms adjust the
difference between the current readings and the known calibration gas
values until they are equal. In this chapter, when we speak about a
calibra-
tion
in a general sense, we are referring to either a verification or a calibra-
tion.
Span and Zero Calibration Gas Definitions
This chapter also refers to the terms
span
and
zero
calibration gases. The
span gas is the
high
calibration gas. The zero gas is the
low
calibration gas.
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...