Serial Communications
|
A-15
29/41
Recov_time
Amount of time for recovery after a
calibration in seconds
IQ
Amount of time for recovery after a
calibration in MMSS
S2
R/W
2A/42
O2_gas1
Oxygen span gas setpoint value, always in
percent
R/W
2B/43
O2_gas2
Oxygen zero gas setpoint value, always in
percent
R/W
2C/44
Co_gas1
Combustibles PPM span gas setpoint value
S2
Combustibles span gas setpoint
IQ
R/W
2D/45
Co_percent_gas
Combustibles percent span gas setpoint
value
S2
R/W
2E/46
Me_percent_gas
Methane percent span gas setpoint value
R/W
2F/47
Cal_cgas1
Oxygen span gas setpoint for last
calibration
Read only
30/48
Cal_span
Oxygen span gas reading for last
calibration
Read only
31/49
Ver_cgas1
Oxygen span gas setpoint for last verify
Read only
32/50
Ver_span
Oxygen span gas reading from last verify
Read only
33/51
O2_span
Oxygen span gas millivolts during last
calibration
Read only
34/52
Cal_cgas2
Oxygen zero gas setpoint value for last
calibration
Read only
35/53
Cal_zero
Oxygen zero gas reading from last
calibration
Read only
36/54
Ver_cgas2
Oxygen setpoint value used during last
verify
Read only
37/55
Ver_zero
Oxygen zero gas read during last verify
Read only
38/56
O2_zero
Zero gas millivolts read during last
calibration
Read only
39/57
Cal_cgas3
Combustibles span gas setpoint during last
calibration
Read only
3A/58
Cmbspan
Combustibles span value during last
calibration
Read only
3B/59
Ver_cgas3
Combustibles span gas setpoint for last
verify
Read only
3C/60
Vcmbspan
Combustibles span gas read during last
verify
Read only
3D/61
Cmbzero
Combustibles zero gas read during last
calibration
Read only
3E/62
Vcmbzero
Combustibles zero gas read during last
verify
Read only
3F/63
Cal_cgas4
Methane span gas setpoint during last
calibration
Read only
40/64
Mthspan
Methane span value read during last
calibration
Read only
41/65
Ver_cgas4
Methane span gas setpoint for last verify
Read only
42/66
Vmthspan
Methane span value during last verify
Read only
43/67
Mthzero
Methane zero value during last calibration
Read only
44/68
Vmthzero
Methane zero value during last verify
Read only
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...