4-22 | Thermox Series 2000 WDG Insitu Analyzer
Alarm Configure
Define whether alarms should trigger based on a high alarm or a low
alarm condition.
Whether Alarm 3 is set high or low only applies if it is set to track
oxygen levels (% O
2
). If Alarm 3 is set to trigger at the start of a cali-
bration or verification, this option does not apply.
Configuring the alarms
1. From the
Alarm
menu, select
Alarm Configure
.
2. Define Alarm 3 and Alarm 4 as high or low alarms.
3. Press
Cancel
to exit this menu option.
Relay Configure
All Series 2000 control unit relays are Form A, normally open contacts.
You can use the software to set whether these alarms should close on
an alarm condition (energize on alarm) or open on an alarm condition
(de-energize on alarm). This menu option applies to all alarms except for
the service and watchdog alarms, which are always set to de-energize on
alarm (fail-safe).
Configuring the relay
1. From the
Alarm
menu, select
Relay Configure
.
2. Select
De-energize on Alarm
or
Energize on Alar
m.
If you select
Energize on Alarm
, the following message appears:
Relays are set to
ENERGIZE on Alarm
If you select
De-energize on Alarm
, the following message appears:
Relays are set to
DE-ENERGIZE on Alarm
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...