Installation and Start-Up | 3-3
sensor Mounting
The ambient temperature in the junction box should be lower than 160°F (70°C).
If you purchased the Division 2 option, the maximum ambient temperature range
is 140°F (60°C). If the ambient temperature is greater than this, consult the fac
-
tory. Mount the probe and its flange to the flange on the process (Figure 3-2). A
length of threaded 3/4” pipe may be temporarily screwed to the probe flange to
support the probe as you insert it into the process. Insulate the pipe nipple with
at least 1-inch thick (2.54 cm) weatherproof insulation. If the pipe nipple is more
than 4 inches long (10.16 cm), it should be heat-traced.
Division 2 option
If you purchased the Division 2 option, follow all warnings as stated on the In-
situ, Division 2 warning label. See Figures 3-3 and 3-4 for mounting dimensions
and plumbing connections.
The Z-Purge apparatus provided by Thermox meets NFPA 496 under
the following conditions.
• You must provide a protective supply with an alarm for a loss of pressure
indication.
• Do not exceed the maximum temperature rating listed on the warning label.
• Process pressure must always be subatmospheric (vacuum).
NOTE
Содержание Thermox WDG Insitu
Страница 12: ...xii Thermox Series 2000 WDG Insitu This page intentionally left blank...
Страница 20: ...1 8 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 24: ...2 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 28: ...3 4 Thermox Series 2000 WDG Insitu Analyzer Figure 3 2 Insitu senor mounting...
Страница 44: ...3 20 Thermox Series 2000 WDG Insitu Analyzer Figure 3 19 Sensor connections...
Страница 51: ...Installation and Start Up 3 27 Figure 3 23 Series 2000 2 Wire RS 485 communications connections...
Страница 53: ...Installation and Start Up 3 29 Figure 3 24 Series 2000 4 wire RS 485 communication connections...
Страница 114: ...6 16 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 128: ...7 14 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 154: ...A 26 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 158: ...B 4 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...
Страница 164: ...D 2 Thermox Series 2000 WDG Insitu Analyzer This page intentionally left blank...